Skip Navigation

  • HOME
RADIO IMMUNOASSAY TUMOR MARKERS

SCC

Immunoradiometric assay for quantitative determination of antigen for squamous cell carcinoma in human serum

Summary

  • KFDA Registration No

    14-3145

  • CAT No

    RC12N

  • TEST METHOD

    IRMA

  • SAMPLE VOLUME

    25 ul

  • INCUBATION TIME

    90'RT

  • STD RANGE

    0-150 ng/ml

Download File

Intended Use

Immunoradiometric assay for quantitative determination of antigen for squamous cell carcinoma in human serum

PRINCIPLE OF THE ASSAY

The RIAKEY SCC IRMA Tube is a non-competitive immunoradiometric (IRMA) method (“sandwich”). The method employs two highly specific SCC antibodies which recognize two different epitopes of the molecule. One antibody is coated on solid phase (coated tube), the other, specific for the SCC and labeled with Iodine-125, is used as a tracer. Antibody-coated polystyrene tubes serve as solid phase. The tracer antibody and the coated antibody react simultaneously with the SCC antigen present in the standards, control serum and samples for reaction. Unbounded material is removed by a washing step. The amount of bound tracer will be directly proportional to the SCC antigen concentration and the remaining radioactivity bound to the tubes is measured in a gamma scintillation counter.

Use Precaution

Be careful when handling all samples, reagents, or devices used in the test as they may be the source of infection.
All reagents, human body samples, etc. are handled at the designated location.

  1. Do not use mixed reagents from different lots.
  2. Do not use reagents beyond the expiration date.
  3. Use distilled water stored in clean container.
  4. Use an individual disposable tip for each sample and reagent, to prevent the possible cross-contamination among the samples.
  5. Store the unused coated tubes at 2~8ºC in the appropriate bags with silica gel and accurately sealed.
  6. If large quantity of assay would be performed at one time, there might be substantial time variation between 60 tubes at one time to minimize time variation. Also, do not exceed 10 minutes for entire pipetting.
  7. Wear disposable globes while handling the kit reagents and wash hands thoroughly afterwards.
  8. Do not pipette by mouth.
  9. Do not smoke, eat or drink in areas where specimens or kit reagents are handle.
  10. Handle samples, reagents and loboratory equipments used for assy with extreme care, as they may potentially contain infectious agents.
  11. When samples or reagents happen to be split, wash carefully with a 3% sodium hypochlorite solution.
  12. Dispose of this cleaning liquid and also such used washing cloth or tissue paper with care, as they may also contain infectious agents.
  13. Avoid microbial contamination when the reagent vial be eventually opend or the contents be handled.
  14. Use only for IN VITRO.